Displaying report 1-1 of 1.
Reports until 14:20, Friday 19 April 2013
H1 SEI
filiberto.clara@LIGO.ORG - posted 14:20, Friday 19 April 2013 (6139)
SEI END Y
Modified Trillium 240 Interface chassis at END Y to D1000749-v3. Resistors R30, R33, R37, and R38 were replaced with 20KOhm (Modification stops U7B from saturating).

Serial Numbers of Units:
S1103178
S1103179
S1103176
Displaying report 1-1 of 1.