New DCPD preamp works. DCPD_B is the one that sees the reflected beam as of now (Daniel, Betsy, Keita)

I used a flashlight (through the OMC breadboard) to aim at the DCPD in reflection of the 50:50 BS as best as I can , and DCPD_B showed much stronger signal.

This means that:

  1. DCPD preamp and whitening chassis work.
  2. DCPD_B is the one that sees reflection as of now, DCPD_A is receivning the transmission.

See cartoon, this is a top view looking through the glass OMC breadboard (optical components are under the breadboard).

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 14:15, Thursday 27 January 2022 (61490)

Wiring diagram is here D1900511-v7 with ISC-R5 on pg 21.