Displaying report 1-1 of 1.
Reports until 15:19, Tuesday 01 February 2022
H1 TCS
camilla.compton@LIGO.ORG - posted 15:19, Tuesday 01 February 2022 (61577)
TCSX VP Enclosure to Table Adapter beam clipping fixed by Tyler!

As TJ showed photo in alog 61382, we were concerned that the TCS CO2 VP Enclosure to Table Adapter Interface Assy. (D1600127) would clip our CO2X beam.

Tyler increased the length of the slots of D1600112 to slide the assembly further in -x and removed some of the internal thickness of D1600109. There is now no risk of clipping as seen in the attached photo- taken though the 2 iris along the beam path. Thank you Tyler! 

I installed a new bellow and attached a photo of the alignment laser on the ZeSe secondary VP. 

Images attached to this report
Displaying report 1-1 of 1.