Reports until 17:11, Thursday 07 April 2022
H1 ISC
elenna.capote@LIGO.ORG - posted 17:11, Thursday 07 April 2022 - last comment - 18:22, Thursday 07 April 2022(62507)
OMC Scans at 10W and 50W input

[Jenne, Sheila, Georgia, Craig]

The IFO was powered up on RF so we could perform OMC scans. The template for the scan is located in userapps omc/h1/templates/OMC_scan_50W.xml

80 second scan duration

Scan GPS Time Darm offset (DARM1_offset) Input Power (W) Circ Power (kW, X arm) RF 9 mod depth (dBm) RF 45 mod depth (dBm)
1 plot 1341004914 0 10 51 23.4 18
2 plot 1341008429 1e-5 10 51 23.4 18
3 plot 1341009831 0 50 233 20.4 18
4, attempted lockloss 0 50 233 20.4 15

Scan 4 did not complete- lost lock at 22:53 UTC. Likely due to 3 dB reduction of RF45 modulation depth (lock held for two minutes after reduction).

Comparison plot of of 10 W and 50 W scan

Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 18:22, Thursday 07 April 2022 (62515)

Just a note that I killed the lock when trying to go from DC readout at high power to RF readout.  I had matched the gains via the input matrix, but we lost lock during the ramp between them.  Next lock Georgia figured things out so that we were on RF DARM.