(Caveat: Since the alignment was not really touched up, alignment fluctuation might be coming into play. Wait for WFS commissioning for a better data.)
Jan Pold asked and here it is.
1 Watt into MC. The transmission is monitored with IM4 transmission QPD as well as MC2 transmission QPD.
The whitening gain for both of the QPDs were 18dB (though I could have set it higher I don't think that would have made any difference).
Three stages of whitening filters were used for both: For MC2, it is a fixed one stage (0.4:40) plus two stages of switchable (1:10), and for IM4 since the fixed stage was not present I used all three switchable stages.
60Hz comb filter was turned off.
The input to each segment before dewhitening was 1000 to 6000 counts (mostly AC), so it's decent.
The beam was not centered at all: For MC2, [P,Y]=[0.14, 0.55], and for IM4 this was [0.22, 0.07].
Anyway, the spectrum was measured for SUM_IN1_DQ channel, divided by DC and plotted. The data was saved as
/ligo/home/controls/keita.kawabe/RIN
ISS status:
Digital loop on, first loop on, second loop off.
PD B was used.
PSL-ISS_GAIN 9.6dB.
Diffraction 13 to 16%.
It turned out that HVAC was/is blasting full. We'll try science mode -ish HVAC with Robert at some stage.