WP10519, 10523 Power Cycle h1susex, replace DAC
Rahul, Erik, Fil, Dave, Tony:
At 09:30 h1susex was powered down. The order was: front end computer, IO Chassis, AA and AI chassis. The power up sequence was AA chassis, IO Chassis, front end then AI chassis.
We found that on restart h1iopsusex was reporting a failed autocal on the first DAC (and 18 bit card). We retested by stopping the models and then running the iop. Again the first 18bit DAC had failed its autocal. Fil took a spare card out and replaced the failed unit.
New Beckhoff Code
Daniel, Dave:
Daniel installed new Beckhoff code this morning, which resulted in modified DAQ INI files for AUX [EX,EY], TCS [EX,EY] and SQZ CS.
DAQ restart
Dave:
The DAQ and the EDC were restarted for the Beckhoff INI changes. h1daqgds1 needed a second restart to sync its channel list.
CER network switch reconfiguration
Fil, Dave:
sw-lvea-aux (which is actually in the CER) was configure to provide three ports to the LVEA serving the workstation network.
Restart/Reboot Log
Tue28Jun2022
LOC TIME HOSTNAME MODEL/REBOOT
09:51:51 h1susex ***REBOOT***
09:53:49 h1susex h1iopsusex
09:53:56 h1susex h1susetmx
09:54:04 h1susex h1sustmsx
09:54:11 h1susex h1susetmxpi
11:29:57 h1susex h1iopsusex
11:56:08 h1susex ***REBOOT***
11:58:09 h1susex h1iopsusex
11:58:16 h1susex h1susetmx
11:58:24 h1susex h1sustmsx
11:58:31 h1susex h1susetmxpi
14:34:18 h1susauxb123 h1edc[[DAQ]]