Reports until 10:49, Tuesday 09 August 2022
H1 TCS
camilla.compton@LIGO.ORG - posted 10:49, Tuesday 09 August 2022 (64385)
ETMY HWS new references but ITMY reflection still getting to camera

After replacing the ETMY HWS plate last week (alog 64287), this morning I took new HWS ETMY references with ALS shuttered and HWS camera at 20Hz. EY ring heater is on at 1W/segment. Sadly the ITMY retro-reflection is still getting back to the EY camera. See images with and without ITMY aligned. More work will need to be done to get rid of this refelction.

Images attached to this report