Displaying report 1-1 of 1.
Reports until 16:26, Tuesday 16 August 2022
H1 ISC
elenna.capote@LIGO.ORG - posted 16:26, Tuesday 16 August 2022 (64523)
PRC2 P filters updated

I updated my modified plant inversion filter that I originally created in alog 64374 to prevent ringing in PRC2 P around 3 Hz. I tested this filter during DRMI acquisition and at low noise and it works very well. I have now updated the ISC_LOCK and ISC_DRMI guardians, and updated sdf to now use FM8 instead of FM7.