Displaying report 1-1 of 1.
Reports until 12:39, Monday 17 October 2022
H1 SQZ
daniel.sigg@LIGO.ORG - posted 12:39, Monday 17 October 2022 (65359)
Laser Locking Revisited

We revisited the squeezer TTFSS with the new FSS settings described in alog 65236 which should fix the 2.5MHz line discovered in alogs 64653 and 64649. With a FSS common gain of 11dB and a fast gain of 8dB the 2.5MHz was still visible but with a 20dB lower amplitude. This allowed us to get back to the old squeezer TTFSS settings and lock with the EOM.

Playing with the FSS common and fast gain, there is a strange effect where lowering the common gain with the old fast gain of 11dB makes the 2.5MHz disappear completely but a peak at 160kHz will appear instead. The later is probably a PZT resonance. Lowering the fast gain will then make the 160kHz peak disappear but the 2.5MHz peak will again appear—albeit at a lower amplitude.

Images attached to this report