J. Kissel for everyone who's asking: We've gathered a bit of "quiet time" when no one was touching the IFO, just after a broadband PCAL injection to test the state of the calibration. Start: 2022-Nov-19 00:35:30 UTC 2022-Nov-18 16:35:30 PST 1352853348 End: 2022-Nov-19 00:45:30 UTC 2022-Nov-18 16:45:30 PST 1352853948 There's far too many folks here for me to summarize the current state of the IFO beyond what I know, but this is supposed to be a time period that we can compare to time prior to the Sep/Oct 2022 vent. - Well thermalized, having been locked for hours - 50.2 W into the IMC, 47.6 W into PRM - DARM drive is all ETMX - The PUM drivers are in State 1 - No Squeezing - Still the same "fudges" to the calibration from LHO:64316 BUT, now - No HAM5/6 Septum window - Some "test" configuration of the EX electrical grounding situation The broadband PCAL injection (again with no updates to the calibration) taken just prior to this quite time lives here: /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/2022-11-19_H1_PCALY2DARMTF_BB.xml