ITMY M0 to M0 and R0 to R0 transfer functions have been running over the weekend for phase 3b testing
The attached documents are showing comparisons between :
Most of the transfer functions are in good agreeement with model and previous measurements. Similarly to Mark's comment for phase 3a testing on aLog 5304, there's a discontinuity at 0.4Hz on the main chain TF in vertical and roll (page 3 and 4 of first pdf), probably due to a measurement artifact. A dtt measurement will be needed to confirm it.
scripts and data have been commited to the svn