ITMY Phase 3b transfer functions

ITMY M0 to M0 and R0 to R0 transfer functions have been running over the weekend for phase 3b testing

The attached documents are showing comparisons between :

Most of the transfer functions are in good agreeement with model and previous measurements. Similarly to Mark's comment for phase 3a testing on aLog 5304, there's a discontinuity at 0.4Hz on the main chain TF in vertical and roll (page 3 and 4 of first pdf), probably due to a measurement artifact. A dtt measurement will be needed to confirm it.

scripts and data have been commited to the svn