Dan, Mitchell, Camilla. ITMX (which has known point absorbers) is absorbing ~30% more power than ITMY.
Measured IFO beam self-heating absorption using this morning's lockoss. Dan left on the the CO2's on after loosing lock (66054) so that they don't change the state of the ITMs. I have now turned both CO2'soff to start locking and Dan's reverted ISC_LOCK.
See attached plot made using Aidan's absorption plot script (instruction in TCS wiki) from in /ligo/gitcommon/labutils/hws_absorption_fit/nov2022.
ITMX: 115mW absorbed = 490ppb
ITMY: 90mW absrobed = 385ppb
with arm power 235kW from H1:ASC-{X/Y}_PWR_CIRC_OUT16. Previously measured 430ppb for ITMX (62782) and 370ppb for ITMY (62468) .
____________
Also tried similar comparison of spherical power during power up after 50W and got same result:
Ring heater test 1350878491 | Power up 25W to 50W* @1353397628 compared to +1h30 (locked with no CO2) | |||
Measured Spherical Power (uD) | Simulated | Measured Spherical Power (uD) |
Simulated (different from self-heating gains) |
|
ITMX | -157 | -67 | 80 | 17 |
ITMY | -150 | -67 | 60 | 23 |