H1 PEM
james.batch@LIGO.ORG - posted 14:53, Tuesday 04 June 2013 (6622)
Restarted h1pemmx
Cyrus, Jim

The h1pemmx models died 12:22:42 UTC June 2.  

At Mid X, found that portable power supply supplying the +24V to the I/O chassis had died, along with a mouse that was laying next to it.  Replaced the portable supply with a Sorenson from the EE Lab (given to us by Filiberto).

Power cycled the timing system hoping to get rid of the 1/2 second time difference that appears on the MEDM screens.  This did not have any visible effect on the timing, so we suspect we may have to wait until we get a known good pair of fibers for timing to the mid station identified before this is cleared up.

Powered up the I/O chassis and front-end computer, the h1pemmx model is now running.