filiberto.clara@LIGO.ORG - posted 11:51, Tuesday 24 January 2023 (66965)
Grounding of EY Electronics in Electronics Bay

WP 10931

A wire braid was used to ground some of the electronics at EY. Same as EX (alog 66402 and alog 66469). Three wire braid cables were pulled from the isolated grounding terminal to the following racks:

  1. ISC-C1
  2. TCS-C1
  3. FAC-C1

Method provides a low resistance path to the terminal block. A list of electronics with grounding modified:

  1. ISC-C1 rack, all chassis except the IO chassis were grounded
  2. TCS-C1 rack, slow controls chassis, slow controls concentrator, and Safety System Controls Interface Chassis were grounded
  3. FAC-C1 rack, Safety System Controls Chassis was grounded