Compare digital AA filter between pydarm production and FOTON export

Hsiang-Yu Huang

Compare digital AA filter between pyDARM production and Foton export (aka Dec65K in H1OMC_DCPD_A0)

I use digital_aa_or_ai_filter_response() in pyDARM to compute this.
I set up the following in sensing part of INI file.
[sensing]
anti_aliasing_rate_string = 512k-daq
anti_aliasing_method      = biquad

Attachment:

Figure. 1 : freq. vs. magnitude(dB) plot by Foton

Figure. 2 : freq. vs. magnitude plot also by Foton

Figure. 3 : compare digital AA filter between pyDARM production and Foton export

Figure. 4 : zoom-in plot from Figure. 3. Set freq. limit from 10 Hz to 2000 Hz in all subplots.  Set up y-axis limit of magntiude residual plot to +- 0.1

               Phase limit to +- 5 degree.

Figure 5 : I export Dec65K in this OMC-DCPD_A0.

Discussion:

For magnitdue part, they are consisent below 1 kHz

However, we have phase loss from ~ 100 Hz. In 1 kHz, there is almost 5 degree phase loss.

Images attached to this report
Comments related to this report
evan.goetz@LIGO.ORG - 00:38, Saturday 11 February 2023 (67350)
The 512k-daq string refers to the 8x DAQ downsampling filter and not the filter module from the FOTON filter file. You should use the omc_filter_file section in the ini file with gain, modules, and bank name to get the right filter coefficients. This is the reason why you see such a large difference. 
hsiang-yu.huang@LIGO.ORG - 02:27, Saturday 11 February 2023 (67351)

Hsaing-Yu Huang

I misunderstand the function I used. I re-calculate the digital AA by omc_digital_filters_response() in pyDARM sensing.

Images attached to this comment