jeffrey.kissel@LIGO.ORG - posted 09:55, Friday 24 February 2023 - last comment - 12:35, Friday 24 February 2023(67600)
Initialized 50-ish models worth of IPC_REMOTE_PAUSE
J. Kissel

I've taken some time this morning to initialize a single channel that has set off the middle yellow flag (H1:FEC-${DCUID}_SDF_UNINIT_CNT) in all front-end model's SDF systems.
The appearance of this channel, H1:${MODELNAME}_REMOTE_IPC_PAUSE, is discussed in LHO:67598.

Process:
    - Changed TABLE SELECTION to CHANS NOT INIT
    - Confirm that the setting value is 0.0
    - Toggled ACCEPT and MON
    - Hit CONFIRM
    - Changed TABLE SECTION back to SETTINGS DIFF

I list below the models I accepted and monitored. 
Models which have the *** next to them show the "hyphen-underscore" bug mentioned in LHO:67598.
EY                                 CORNER                             EX                              FCES
model       snap file              model       snap file              model        snap file          model          snap file
h1susetmy   safe                   h1asc***    safe                   h1susetmxpi  safe               h1sqzfces      safe
h1sustmsy   safe                   h1ascimc    safe                   h1susetmx    safe               h1susauxh8     safe
h1isietmy   OBSERVE                h1lsc***    safe                   h1sustmsy    safe
h1hpietmy   OBSERVE                h1lscaux    safe                   h1isietmx    OBSERVE
h1iscey     safe                   h1omc***    safe                   h1hpietmx    OBSERVE
h1alsey     safe                   h1ascsqzifo safe                   h1iscex      safe
h1pemey     safe                   h1ascsqzfc  safe                   h1alsex      safe
h1etmypi    safe                   h1omcpi     safe                   h1pemex      safe
                                   h1susprocpi safe
                                   h1susitmpi  safe  
                                   h1oaf***    safe
                                   h1tcscs     safe
                                   h1pemcs     safe
                                   h1susproc   safe
                                   h1seiproc   safe
                                   h1susauxh7  safe
                                   h1psliss    safe
                                   h1psliss    safe
                                   h1pslpmc    safe
                                   h1psldbb    safe

There still remains all the corner station SUS and SEI models, but I need to move on to other things. 
Here's hoping the operator core has some spare time to follow my instructions for the remaining models, or they or the CDS team can write a script to quickly accept and monitor these channels.
Comments related to this report
ryan.crouch@LIGO.ORG - 12:35, Friday 24 February 2023 (67602)

I have gone through the remaining corner station sus and sei models and accepted and monitored for the following

Corner:

h1hpiham1       observe       h1susmc1     safe       h1susmc3        safe        h1susprm      safe       h1suspr3      safe      h1isiham2     observe

h1hpiham2       observe       h1susim       safe       h1sushtts       safe        h1susmc       safe       h1suspr2      safe      h1isiham3      observe

h1hpiham3       observe       h1sussr2      safe        h1isiham4      observe     h1hpiham4     observe    h1sussr3       safe      h1sussrm       safe

h1isiham5       observe       h1phiham5    observe     h1susifoout     safe        h1sussqzout    safe       h1isiham6      observe   h1hpiham6    observe

h1susfc1        safe          h1sussqzin    safe        h1isiham7      observe     h1pemh8      safe       h1susfc2       safe       h1isiham8     observe

h1susitmy      safe           h1isibitmy    observe     h1susbs        safe        h1isibs        observe    h1susitmx      safe       h1isiitmx      observe


 
Displaying report 1-1 of 1.