Lock 1:
Lock 2:
3 hour mark ASC issues:
Bruco from 10 minutes quiet time: https://ldas-jobs.ligo-wa.caltech.edu/~elenna.capote/brucos/OAF_1362267774/
Just a potential clue, since there were a succession of few-hour locks last night, each with a slow drop in range — the common mode wavefront sensors show a similarly slow drift, especially in yaw. This is not necessarily surprising, significant, or new. Since you control six DOFs (cHard, PR2, IM4 pitch/yaw) with linear combinations of eight sensor channels (REFL A/B RF9/45 I pitch/yaw), there are two remaining linear combinations of sensors that are not servoed. The question is whether the drift actually means anything. It could just be harmless thermalization, if such a thing exists. Or it could be (for example) an indication that the PRM alignment drifts away since it is sensed from a spot dither rather than wavefront sensing. An SRC optic could also be the culprit.