Displaying report 1-1 of 1.
Reports until 18:11, Monday 27 March 2023
H1 CAL
daniel.brown@LIGO.ORG - posted 18:11, Monday 27 March 2023 - last comment - 13:56, Tuesday 28 March 2023(68179)
Manual calibration update

We updated the calibration by hand, adjusting H1:CAL-CS_DARM_ERR_GAIN => 0.93 which can be seen in the jump in range.

Images attached to this report
Comments related to this report
louis.dartez@LIGO.ORG - 11:54, Tuesday 28 March 2023 (68196)
I updated H1:CAL-CS_DARM_DELTAL_RESIDUAL_WHITEN_GAIN to 1/0.93 to compensate for the 0.93 fudge in H1:CAL-CS_DARM_ERR_GAIN in the path that goes to GDS.

GPS time: 1364064564.9382627
Images attached to this comment
jenne.driggers@LIGO.ORG - 13:56, Tuesday 28 March 2023 (68207)

It seems that we had forgotten to SDF these, so they were reverted when we started locking today.  I have now put the values back in, and saved them in SDF.  See attached for the values I put in.

Images attached to this comment
Displaying report 1-1 of 1.