H1 TCS
camilla.compton@LIGO.ORG - posted 10:01, Monday 03 April 2023 (68356)
No evidence of water leak on CO2X table.

Closes WP 11090. We've been having to add a large amount of water to the CO2X chiller ( alog 68277), > 4L in 2 weeks. I checked on the CO2 table and saw no water on the table or sign of a new leak. 

There is some potential new water marks near the RF driver (back right +X corner of table). See attached photos, but these could be old as we've had a water leak on this table before. Nothing was wet, no drips. Fil and I did touch the RF driver cables on March 7th which is the day Austin found we had to add extra water. 

We can try to walk the lines again tomorrow during maintenance. 