J. Kissel, H-y. Huang, L. Dartez While we were perusing and updating the pyDARM parameter file (which lives in pydarm_H1.ini), we were reminded of the error in whitening chassis compensation of the OMC DCPD B channel of that I (Jeff) installed back in Feb (chassis install LHO:67110), namely, when I installed Louis' fits, from LHO:67114, I errantly copied DCPDA's poles and zeros into both the DCPDA and DCPDB "NewAW" filter banks. We've now fixed this error in DCPD B compensation, and while we were at it, we increased the precision in the DCPD A compensation. As such, now, the "NewAW" filters are as they should be based on the "whitening ON" fit results: Filter Bank Module Name Design String OMC_DCPD_A0 FM2 "NewAW" zpk([9.902126],[0.996556],1,"n") OMC_DCPD_B0 FM2 "NewAW" zpk([9.88272],[0.994856],1,"n") This change has been loaded in and running as of 2023-04-07 19:15 UTC, and the filter file has been committed to the userapps repo under /opt/rtcds/userapps/release/cds/h1/filterfiles/H1IOPOMC0.txt. Note -- this means we'll need to redo the change in the OMC DCPD SUM and NULL balance matrix, LHO:68403. *sigh*