H1 ISC
daniel.sigg@LIGO.ORG - posted 18:36, Friday 21 June 2013 (6852)
Green ALS PD in transmission recalibrated
The calibration of the green ALS PD in transmission of the arm cavity and located on ISCT1 have been remeasured. The responsitivity of the PD was changed to 0.22 A/W (from 0.755 A/W); a factor of 3.4 decrease. The beamsplitter ratio was measured to be 19% (from a theoretical 20%). The datasheet of the SM1PD1A indicated around 0.23 A/W at 532 nm.