FAMIS 18393 and 23737 (as I was late to testing earlier this year)
DM5 - HAM2: Has been turned off for O4 as it is not a pump-less monitor
DM6 - HAM6: Failed both zero-count and flow rate tests; has since been replaced (alog 68592)
DM10 - BSC1: Flow rate 2.8 lpm, passed zero-count test on second try
PSL101 - Laser Room: Flow rate 2.8 lpm, passed zero-count test
PSL102 - Anteroom: Flow rate 2.8 lpm, passed zero-count test
HH1: Makes startup/shutdown noises, but screen does not turn on
HH2: Flow rate 2.8 lpm, passed zero-count test (handheld has a VERY short battery life)
HH3: Flow rate 2.8 lpm, passed zero-count test
HH4: Flow rate 2.8 lmp, passed zero-count test
HH5: Flow rate 2.8 lpm, passed zero-count test
Diode Room: Flow rate 2.8 lpm, passed zero-count test
Optics Lab: Flow rate 2.8 lpm, passed zero-count test
Vac Prep Lab: Flow rate 2.8 lpm, passed zero-count test
EX - VEA: Flow rate 2.7 lpm, passed zero-count test on second try
EY - VEA: Flow rate 2.8 lpm. passed zero-count test
FCES - HAM8: Flow rate 2.5 lpm, passed zero-count test (I was told after this test that the flow rate can be adjusted; I did not do this, but the monitor was running stably, so I don't expect this needs replacing)