Displaying report 1-1 of 1.
Reports until 09:56, Thursday 27 April 2023
H1 SUS (GRD, OpsInfo, SUS)
rahul.kumar@LIGO.ORG - posted 09:56, Thursday 27 April 2023 - last comment - 10:13, Thursday 27 April 2023(69080)
SUS SDF accepted for OBSERVE STATE

Betsy, TJ, Rahul

I have accepted or unmonitored a bunch of settings on SUS SDF for the OBSERVE STATE. I trended all the settings difference and accepted everything which didn't change in the last few months. I have also set the offsets (for COILOUTF) to zero for all the QUADs and R0 chain after going through all of them. There were a few things which were changed by the GRD, hence I have either accepted them or unmonitored (things like TRAMP) them.

Attached below are the screenshot of all the SDF which original existed when I started yesterday morning. Hence if changes were accepted by mistake then please compare it with the screenshot and make the necessary changes.

Still there are a few SUS SDF remaining and we are planning to tackle this today (currently ongoing).

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 10:13, Thursday 27 April 2023 (69085)

All of the remaining SUS observe.snap SDFs are now cleared. I checked the remaining and tracked them down to either exsisting in the safe.snap at the accepted value, or coded into the guardian at that value.

Images attached to this comment
Displaying report 1-1 of 1.