J. Betzwieser, D. Bhattacharjee, J. Kissel, R. Savage Executive summary: More changes to calibration lines in order to commission the best answer out of the PCALXY comparison. Based on the systems level compromises discussed on Friday last week about having the PCALX contribution to the PCALXY comparison right next to the 410.3 Hz PCALY line that's used for TDCFs (LHO:69175), I moved the PCALX comparison line from 410.2 Hz (0.1 Hz below 410.3 Hz) to 409.8 Hz (0.5 Hz below 410.3 Hz). LHO:69303. After looking at the results from that move to 409.8 Hz, Joe, Rick, and Dripta re-discovered that there's a notch in the DARM loop at 410.3 Hz (see LHO:48530), which is a relatively wide elliptic bandstop (whose boundaries are 409.4 and 411.2 Hz, i.e. 410.3 +/- 0.9 Hz). That means the frequency response change of IFO's response function, C / (1 + G), is changing quite rapidly between the two 410.3 and 409.8 Hz frequencies. *That* means that the traditional, easy, side-by-side PCALXY comparison of lines to arrive at a ratio of amplitudes -- which assumes that the response function ratio isn't changing between frequencies -- becomes harder, and loop model dependent. As such, yesterday, Joe, Rick, and Dripta got together and posted the proposal outlined last night LHO:69331, to move the lines that inform the PCALXY comparison entirely away from 410.3 Hz, such that the comparison is between H1:CAL-PCALX_PCALOSC8_OSC_FREQ 283.91 Already present as a "Systematic Error" line, but now louder, increasing amplitude from 200 to 2000 cts. H1:CAL-PCALY_PCALOSC4_OSC_FREQ 284.01 New PCALY line dedicated to this test, set with an amplitude of 1430 cts I've implemented their proposal at of 2023-05-05 20:00 UTC. Knock-on effects: - Because there's a new line on PCALY, I wanted to check whether we still have enough OFS range to drive the temporary, ER15 "thermalization" lines, driven by the CAL_AWG_LINES guardian. We do, so we're good there. - The new addition of 284.01 Hz PCALY *may* impact (bias) the demodulation (answer) of the 283.91 Hz PCALX systematic error line, given that that line's front-end DEMOD still uses a +/- 0.1 Hz wide SIG band pass, and 0.1 Hz corner frequency low pass, i.e. a 10 second FFT. - Because the systematic Error line is much louder, and has a new companion, equally loud "next door," the CW searches may be more likely to complain about sidebands and non-linearities not subtracted out. - Because I'm using a new PCALY oscillator, I need to double check that the GDS pipeline still safely subtracts this out in the GDS-CALIB_STRAIN_NOLINES channel. - This renders the DEMOD10 PCALX non-functional, since it's stuck demodulating the 409.8 Hz PCALX line that isn't there. Ideally, we'd convert PCALX DEMOD 10 to watching the new 284.01 Hz PCALY line, but that would require a front-end code change. Eventually, once we're done the 24.5 Hz PCALY thermalization line, then we can switch that PCALY DEMOD 4 over to using that. Things that need to change in order to accommodate this move: (1) Accepted new oscillator settings in h1calex and h1caley SDF systems. (2) Modified the pydarm_H1.ini parameter file which now is up to version marked by git hash 156230c7. Parameters cal_line_cmp_pcalx_frequency and cal_line_cmp_pcaly_frequency have changed. (3) Pushed new DARM_ERR / PCAL and DELTAL / PCAL EPICs records at these line frequencies with the updated parameter file, using the command $ pydarm export --push --epics-only --model /ligo/groups/cal/H1/ifo/pydarm_H1.ini (4) Accepted them in h1calcs SDF. Rick said he's going to take care of modifying the DEMOD SIG and low pass filters for the X Y comparisons as he needs. Here's the latest list of calibration lines: Freq (Hz) Actuator Purpose Channel that defines Freq Since O3 15.6 ETMX UIM (L1) SUS \kappa_UIM excitation H1:SUS-ETMY_L1_CAL_LINE_FREQ Amplitude Change on Apr 2023 (LHO:68289) 16.4 ETMX PUM (L2) SUS \kappa_PUM excitation H1:SUS-ETMY_L2_CAL_LINE_FREQ Amplitude Change on Apr 2023 (LHO:68289) 17.1 PCALY actuator kappa reference H1:CAL-PCALY_PCALOSC1_OSC_FREQ Amplitude Change on Apr 2023 (LHO:68289) 17.6 ETMX TST (L3) SUS \kappa_TST excitation H1:SUS-ETMY_L3_CAL_LINE_FREQ Amplitude Change on Apr 2023 (LHO:68289) 33.43 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC4_OSC_FREQ New since Jul 2022 (LHO:64214, LHO:66268) 53.67 | | H1:CAL-PCALX_PCALOSC5_OSC_FREQ Frequency Change on Apr 2023 (LHO:68289) 77.73 | | H1:CAL-PCALX_PCALOSC6_OSC_FREQ New since Jul 2022 (LHO:64214, LHO:66268) 102.13 | | H1:CAL-PCALX_PCALOSC7_OSC_FREQ | 283.91 V V H1:CAL-PCALX_PCALOSC8_OSC_FREQ V 284.01 PCALY PCALXY comparison H1:CAL-PCALY_PCALOSC4_OSC_FREQ TURNED ON THIS ALOG ---- PCALX PCALXY comparison H1:CAL-PCALX_PCALOSC2_OSC_FREQ TURNED OFF THIS ALOG 410.3 PCALY f_cc and kappa_C H1:CAL-PCALY_PCALOSC2_OSC_FREQ No Change 1083.7 PCALY f_cc and kappa_C monitor H1:CAL-PCALY_PCALOSC3_OSC_FREQ No Change n*500+1.3 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC1_OSC_FREQ No Change (n=[2,3,4,5,6,7,8])
Accepted the new value of 284.01 for H1:CAL-CS_TDEP_PCAL_Y_COMPARE_COMPARISON_OSC_FREQ in sdf, screenshot attached.