Reports until 20:37, Friday 05 May 2023
H1 ISC (CAL, ISC, SQZ)
jennifer.wright@LIGO.ORG - posted 20:37, Friday 05 May 2023 - last comment - 10:11, Monday 08 May 2023(69368)
Checking the effect of DARM offset change above 500Hz

Jennie, Naoki, Jenne, Elenna, Vicky

We took more measurements of 40mA vs. 20mA on the OMC SUM DCPD (that is, the DARM offset).

This is because our previous measurements did not correct DARM by the calibration at high frequencies, as the PCAL > DARM broadband transfer functions had low coherence at these frequencies.

The technique above 500Hz is thus to take a DARM ASD spectra and correct it by the peak height of the 1083 Hz PCAL line.

Measurement 1:

21:16:43 UTC

20mA (in OBS), so squeezing injected.

Ref0

 

Measurement 2:

21:38:42 UTC

20mA (IN COMMISSIONING), no squeezing.

Ref1

 

Measurement 3:

21:52:24 UTC

40 mA (IN COMMISSIONING), no squeezing.

Ref2

 

Measurement 4:

22:08:09 UTC

40mA (IN COMMISSIONING), squeezing injected.

Ref3

The measurements are saved in /ligo/home/jennifer.wright/git/DARM_offset/2023-05-05_2114UTC_H1_DARMSPEC.xml.

Attached are the scaled spectra (courtesy of Jenne).

Since the noise looks lower between 40mA no squeeze(brown) and 20mA no squeeze (green), we will go with this offset.
 

We are not sure why the 40mA, squeezer on trace looks so bad at low frequencies. There's not a way for the homodyne angle to produce the given change in the squeezing blue/pink traces, so that is likely the effect of something else on the squeezing. The most relevant comparison should be the no-squeezed DARM traces at different homodyne angles, green vs. brown. 

Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 10:11, Monday 08 May 2023 (69399)

The pink trace, particularly when you look all the way to low frequency (not shown in this plot), includes a glitch time in it.  Not a big deal, since the thing we were looking at here was a lot of the high frequency portion, where the glitch does not seem to be present.