Referencing Camilla's alog 69333, I ran a calib suite.
- Took us out of OBSERVING & took ISC_LOCK to NLN_CAL_MEAS
- Opened terminal and entered/ran: "pydarm measure --run-headless" at 16:40 UTC.
- Script finished at 18:23 UTC
- Moving on to SCRL detuning measurement
Attached is a screenshot of the sitemap > CAL CS > Calibration Monitor screen shortly after starting the measurement.
- Measurement at: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20230508T164115Z.xml
'pydarm report' command gave attached report:
- /ligo/groups/cal/H1/reports/20230508T180014Z/H1_calibration_report_20230508T180014Z.pdf