We ran the functionality test on the main turbopump at EY. The scroll pump is started to take pressure down to low 10^-02 Torr, at which time the turbo pump is started, the system reaches low 10^-08 Torr after a few minutes, then the turbo pump system is left ON for about 1 hour, after the hour the system goes through a shut down sequence.
No issues were encountered while performing the functionality test on this station.
EY Turbo:
Bearing Life:100%
Scroll Pump Hours: 198.8
Turbopump hours: 1266
Closes WP 11179 FAMIS 24935