jeffrey.kissel@LIGO.ORG - posted 13:19, Tuesday 09 May 2023 (69445)
PCALX and PCALY OFS Loops Toggled (Needed reset after Beckhoff Restarts)
J. Kissel, T. Shaffer

While recovering the IFO, we noticed that the ASDs of the PCALX and PCALY on the front wall DARM_FOM looked odd and DIAG_MAIN reported issues with PCALY optical follower servo (OFS). Hearing that there was Beckhoff restarts at the end-stations today, we checked in with the PCAL screens. Indeed, a standard toggle of the OFS loop enable buttons (H1:CAL-PCAL{X,Y}_OPTICALFOLLOWERSERVOENABLE) needed after beckhoff work cleared the issue and the usual spectrum of calibration lines came back up as expected.