H1 General
anthony.sanchez@LIGO.ORG - posted 21:14, Tuesday 09 May 2023 - last comment - 09:42, Wednesday 10 May 2023(69460)
Tuesday Mid EVE shift Update

Locking Went well.
Violin mode ITMY Mode8 was ringing up and I turning off damping for it at 1:15 UTC
Seems to have stopped the mode from ringing up.

Lockloss:
Lockloss from unknown cause at 1:40 UTC:
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1367717996

Relocking
ISC_LOCK Guardian went into and through AQUIRE_PRMI without any intervention and the IFO locked at NOMINAL_LOWNOISE at 2:57 UTC.
The IFO is almost thermalizing and, once it's circulating arm powers settle I'll start a calibration measurement. 
I have attached a picture of the SDF that I have accepted into SDF so i can get back to Observe mode.

CAL_AWG_LINES : ERROR while trying to get to NLN_CAL_MEAS
Reloaded CAL_AWG_LINES Guardian and the problem was resolved.

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 08:51, Wednesday 10 May 2023 (69469)

Picomotor controllers should always be off in observe, i.e. PICO_x_CURRENT_ENABLE should be set to disabled.

camilla.compton@LIGO.ORG - 09:42, Wednesday 10 May 2023 (69471)CAL

I disabled the picomotor that was left on (sitemap > LSC > Picomotors > SQZT 0 > Disabled) and accepted in SAFE sdf. Tagging CAL for the sdf diffs in main alog.