We ran the functionality test on the main turbopump at EX, MX and MY. The scroll pump is started to take pressure down to low 10^-02 Torr, at which time the turbo pump is started, the system reaches low 10^-08 Torr after a few minutes, then the turbo pump system is left ON for about 1 hour, after the hour the system goes through a shut down sequence.
No issues were encountered while performing the functionality test on this station.
EX Turbo:
Bearing Life:100%
Scroll Pump Hours: 6304
Turbopump hours: 263
MX Turbo:
Bearing Life:100%
Scroll Pump Hours: 197
Turbopump hours: 107
MY Turbo:
Bearing Life:100%
Scroll Pump Hours: 11388
Turbopump hours: 198
Closes WP 11198 FAMIS 24911 24863 24839