Updated the TwinCAT code and removed several unnecessary checks from the error handlers:
- LO PFD (H1:SQZ-LO_PFD; no longer installed),
- Beam diverter C (H1:SYS-MOTION_C_BDIV_C; no longer used),
- TCS laser sources (H1:TCS-ETMX_SOURCE and H1:TCS-ETMY_SOURCE; not yet installed),
- New weather stations (H1:PEM-ENDX and H1:PEM-ENDY; not yet installed), and
- DC synchronization (PTP timing)
The EtherCAT overview screen is now mostly green.
3 of the baffle PDs are saturated: H1:AOS-ETMX_BAFFLEPD_1, H1:AOS-ETMY_BAFFLEPD_1, and H1:AOS-ETMY_BAFFLEPD_4. This could be fixed by reducing their gain by 20dB to 0dB, when we are in high power.
The other channel with an issue is H1:PEM-Y_EBAY_RACK1_TEMPERATURE which reads unrealistic values and needs fixing.