Displaying report 1-1 of 1.
Reports until 14:40, Tuesday 16 May 2023
H1 PEM (CDS, DAQ)
david.barker@LIGO.ORG - posted 14:40, Tuesday 16 May 2023 - last comment - 14:59, Tuesday 16 May 2023(69654)
h1pemcs model, channel name changes

Anamaria, TJ, Jonathan, Dave:

A new h1calcs model was installed. Two fast and two slow DAQ channels were renamed.

++: slow channel H1:PEM-CS_ACC_SQZT0_SQZLASER_X_MON added to the DAQ
++: slow channel H1:PEM-CS_ACC_SQZT7_HOMODYNE_X_MON added to the DAQ
++: fast channel H1:PEM-CS_ACC_SQZT0_SQZLASER_X_DQ added to the DAQ
++: fast channel H1:PEM-CS_ACC_SQZT7_HOMODYNE_X_DQ added to the DAQ
--: slow channel H1:PEM-CS_ACC_ISCT6_SQZLASER_X_MON removed from DAQ 
--: slow channel H1:PEM-CS_ACC_SQZT6_HOMODYNE_X_MON removed from DAQ 
-!: fast channel H1:PEM-CS_ACC_ISCT6_SQZLASER_X_DQ removed from DAQ  ***GDS-CHAN***
-!: fast channel H1:PEM-CS_ACC_SQZT6_HOMODYNE_X_DQ removed from DAQ  ***GDS-CHAN***

Total number of DAQ changes = 8
(4 additions, 4 deletions)
 

Note that the fast channels are in the GDS broadcaster channel list. I initially made the change by hand for h1daqgds[0,1] and Jonathan then made the changes permanent in puppet.

h1daqgds0 daqd was down for an extended period while I made the first hand-edit.

Comments related to this report
anamaria.effler@LIGO.ORG - 14:59, Tuesday 16 May 2023 (69658)DetChar

These channels were remnants of the pem changes from O3 to O4 (there is no more SQZT6 or ISCT6), and we overlooked changing them earlier. As such, people who use static channel lists might run into issues with the disappearance of the ISCT6 and SQZT6 channels.
Tagging DetChar.

Displaying report 1-1 of 1.