Took H1 out of Observing to run Calibration Suite Measurements.
Attached is a screenshot (#1) of the sitemap > CAL CS > Calibration Monitor screen shortly after starting the measurement.
'pydarm report' command gave attached report:
Since, PEM team was going to immediately take H1, did the following:
NOTE: CAL_AWG_LINES Goes Into Error (see attached screenshot [#3])
While getting this measurement going, the guardian CAL_AWG_LINES went into ERROR (node is RED).
(I missed this while setting up measurement and then starting draft of this alog. TJ noticed it when he walked in. I should have noticed it from Verbal or from the notification messages on various Guardian windows.)
Jeff mentions that this can happen after there has been a DAQ Restart (there were atleast two yesterday). To get the node back, Jeff hit the LOAD button on CAL_AWG_LINES.