Late entry, tests were done last Tuesday, May 16th 2023.
A functionality test was done on the corner station turbo pumps The scroll pump was started to take pressure down to low 10^-02 Torr, at which time the turbo pump was started, the system reached low 10^-08 Torr after a few minutes, the turbo pump system was left ON for about 1 hour. After the hour the system went through a shut down sequence, turbo is not spinning and power is turned off. No issues were encountered while performing the functionality test on the 3 turbo pump stations.
Findings below:
Output mode cleaner tube turbo station;
Scroll pump hours: 5554.0
Turbo pump hours: 5598.2
Crash bearing life is at 100%
X beam manifold turbo station;
Scroll pump hours: 776.0
Turbo pump hours: 772.8
Crash bearing life is at 100%
Y beam manifold turbo station;
Scroll pump hours: 1870.9
Turbo pump hours: 590.1
Crash bearing life is at 100%