richard.savage@LIGO.ORG - posted 11:52, Tuesday 23 May 2023 (69834)
Quick look at Pcal X/Y calibration comparison factor, chi_XY

Using DTT with 0.001 Hz BW and 10 averages (.xml template attached), chi_XY (see LIGO-P2000113 for details) was calculated by hand during several quiet lock stretches of at least 2 hours during the past few days.

chi_XY values calculated by hand using DTT
Date Start Time chi_XY
5/20/2023 08:00 1.00056
5/20/2023 09:00 1.00079
5/22/2023 11:10 1.00057
5/23/2023 21:00 1.00012

 

 

 

 

 

 

Our calculation of chi_XY in the front-end code needs some tuning of the demodulation and filtering parameters as can be seen in the attached plot.  This will get some attention soon; it is the focus of a planned summer SURF student investigation here at LHO.

chi_XY compares the calibrations of the fiducial Pcal displacements induced at both end stations.  Corrections factors have been applied at each end station (H1:CAL-PCALX_XY_COMPARE_CORR_FACT = 0.9988 and H1:CAL-PCALY_XY_COMPARE_CORR_FACT = 1.0015) such that chi_XY should be close to unity.  It appears that both Pcal systems are stable, operating as expected.