Violin damping phases changed IY5 IY6 rising.

As Ryan C notes in alog 70003, "IX12 & IY8 damping have changed phase, a negative gain damped IX12 down (-2, max of -4) instead of the +2 its set to in lscparams. IY8 also was rising, I changed its gain sign from 0.08 to -0.08 and it started to fall".

Both these modes have been falling well since but our troublesome ITMY Mode5/6 pair have been slowly rising. See attached. I first suspected the IY8 phase change could be causing this to turned the IY8 damping off for 20 minutes. IY8 stopped damping but IY5/6 continued to rise. I've now turned off the mode5 damping, will see what effect this has as I'm worried about adjusting it's phase.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 18:17, Tuesday 30 May 2023 (70022)

ITMX: On Sunday 28th, Tony notes IX12 is ringing up 69972.  Since then, operators have changed gain from +2 to -2 by hand (69973, 69994, 70003). I've updated this in lscparams but cannot reload violin damping until we are out of observing.

ITMY: On Monday 29th, Tony notes IY5, IY6, IY8 are ringing up 69991. He turned off the IY5 gain as I did, this stabilizes them but doesn't damp the modes. I've added this into lscparams to avoid ringing these up until we find damping that works. Plot of IY5/6 today attached.  Tony turned IY8 gain to 0 but Ryan C found that -0.08 damps well so I'll add that to lscparams too.

We lost lock and I reloaded VIOLIN_DAMPING guardian so these lscparams changes (IX12 +2 to -2; IY5 +0.01 to 0, IY8 +0.08 to -0.08) will be in the guardian.

Images attached to this comment