I created new models for the X-arm: h1susitmx, h1susex, h1susetmx and h1sustmsx, liaising with Vincent to get some SEI IPC signals defined. All four models have been successfully compiled and installed. They will not be started until tomorrow. Details are as follows.
h1susitmx:
-
Cloned h1susitmy
-
Changed dcuid to 29 (from 30)
-
Changed specific_cpu to 2 (from 3)
-
Changed comments to give correct wiring diagram, D1100022-v9
-
Changed PCIe inputs and associated EPIC outputs to "...ITMX..."
-
Changed main QUAD block to "ITMX"
-
Left RFM dummied out but changed two error signals to "...ITMX..."
-
Copy'n'pasted all OSEM and OPLEV input signals from l1susitmx (and checked against D1100022-v9)
-
Corrected comment about analog input order for H1:ITMx (from H2:ITMy)
-
Copy'n'pasted all OSEM and OPLEV output signals from l1susitmx (and checked against D1100022-v9)
-
Corrected comment about analog output order for H1:ITMx (from H2:ITMy)
-
Changed PCIe output stuff to "...ITMX..."
-
Grounded TEST1_In and TEST2_In of ITMX block (had been wired to adc_0_29 and adc_0_30)
-
Changed DIO_0 and DIO_1 parts to 2 each of CBO_L32_C0 and CBO_H32_C0 per l1susitmx
-
Compiled/installed OK.
h1susauxex:
-
Cloned h1susauxex
-
Changed all EY to EX, ey to ex, TMSY to TMSX, tmsy to tmsx, ETMY to ETMX and etmy to etmx with text editor
-
Changed dcuid to 114 (from 116)
-
Checked ADC and DAC connections against D1100022-v9
-
Compiled/installed OK.
h1susetmx:
-
Cloned h1susetmy
-
Changed all ETMY to ETMX and all etmy to etmx with text editor
-
Changed host to h1susex (from h1susey) and dcuid to 88 (from 98)
-
Changed DIO_0 and DIO_1 parts to 2 each of CBO_L32_C0 and CBO_H32_C0 per l1susetmx
-
Changed comments to give correct wiring diagram, D1100022-v9
-
Checked ADC and DAC connections against D1100022-v9
-
Removed receiver for IPC signal H1:LSC-ETMX_L_SUSETMX, which is being generated as a PCIE signal by h1lsc (whereas H1:LSC-ETMY_L_SUSETMY is correctly RFM).
-
Compiled/installed OK.
h1sustmsx:
-
Cloned h1sustmsy.mdl
-
Changed all TMSY to TMSX, tmsy to tmsx, ETMY to ETMX and etmy to etmx with text editor
-
Changed host to h1susex (from h1susey) and dcuid to 89 (from 99)
-
Changed DIO_0 and DIO_1 parts to 2 each of CBO_L32_C0 and CBO_H32_C0 per l1sustmsx
-
Compile failed due to IPC channels not yet created on SEI side - same channels as for h1susetmx.
-
Compiled/installed OK.