Following the "by hand" calculation method described in this aLog entry, the X/Y Pcal calibration comparison factor from the 9 hour locked stretch starting at 22:00 UTC on June 5, 2023 is 0.99967 using 10 averages (1024 sec with 50% overlap).
With the same settings and start time, but 60 everages, the calculated X/Y comparison factor is 0.99972
Ideally, it would be unity. This result indicates that it is within 0.03 % of the expected value.