Reports until 00:05, Thursday 08 June 2023
H1 General (ISC)
camilla.compton@LIGO.ORG - posted 00:05, Thursday 08 June 2023 - last comment - 06:00, Thursday 08 June 2023(70267)
OPS Owl Shift Start

TITLE: 06/08 Owl Shift: 07:00-15:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Lock Acquisition
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 4mph 5min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.08 μm/s
QUICK SUMMARY: Just lost lock at 06:53UTC 1370242452 after 2h49 in NLN. Ryan and I watched CSOFT P ASC control signal ring up before the lockloss, plot attached. Tagging ISC.

VAC, SUS, SEI, CDS, dust monitors all Okay.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 06:00, Thursday 08 June 2023 (70275)ISC

Looked over the last locklosses. In general the CSOFT_P control signal does get larger throughout the lock. In particular the last three "unknown" locklosses show CSOFF_P_OUT16 getting up to 4000 in towards the end: this one along with 20230608 0245UTC and 20230608 0653 UTC.

20230607 1635 UTC shows some CSOFT P noise but not as bad. The other locklosses in the last 3 days CSOFT_P_OUT16 stayed below ~2000.

Plot of last 3 days attached.  Did something change during at the "1 day ago" 06/07 09:37UTC? This was the second lock after maintenance and I moved PR3 to get locked 70217.

Images attached to this comment