Displaying report 1-1 of 1.
Reports until 16:36, Friday 09 June 2023
H1 CDS
david.barker@LIGO.ORG - posted 16:36, Friday 09 June 2023 - last comment - 13:01, Tuesday 13 June 2023(70317)
unresponsive filters not under local model control

Following up on EJ's alog, I have extended the code to check if the unresponsive filter is under local control by the model. I ran it this afternoon while H1 was in observe, running in 'nice' mode so as to not hammer the frontends with CA requests (delays between requests).

If we disregard locally controlled filters, the number of unresponsive filters is reduced to 3 (see attached). The top one runs on h1lsc, the other two on h1lscaux.

Here is the full list, with locally controlled filters marked with *

Total numer of filtermodules = 13367
num unresponsive 14

h1lsc {'LSC-EXTRA_AI_2': 'FM2  '}

h1lscaux {'LSC-LOCKIN_1_DEMOD_9_I': 'FM1  ',
          'LSC-LOCKIN_1_DEMOD_9_Q': 'FM1  '}

h1sqz {'SQZ-RLF_VCXO_SERVO': 'FM2* '}

h1alsex {'LSC-X_ARM_DRIVE': 'FM3* FM5* '}

h1alsey {'LSC-Y_ARM_DRIVE': 'FM3* FM5* '}

h1sussqzin {'SUS-ZM1_M2_COILOUTF_LL': 'FM1* FM6* ',
            'SUS-ZM1_M2_COILOUTF_LR': 'FM1* FM6* ',
            'SUS-ZM1_M2_COILOUTF_UL': 'FM1* FM6* ',
            'SUS-ZM1_M2_COILOUTF_UR': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_LL': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_LR': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_UL': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_UR': 'FM1* FM6* '}
 

 

 

 

 

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 14:38, Monday 12 June 2023 (70378)

Turned thses filters off:

h1lsc {'LSC-EXTRA_AI_2': 'FM2  '}
h1lscaux {'LSC-LOCKIN_1_DEMOD_9_I': 'FM1  ',
          'LSC-LOCKIN_1_DEMOD_9_Q': 'FM1  '}

Added gain of 1 filters to:

h1sqz {'SQZ-RLF_VCXO_SERVO': 'FM2* '}
h1sussqzin {'SUS-ZM1_M2_COILOUTF_LL': 'FM1* FM6* ',
            'SUS-ZM1_M2_COILOUTF_LR': 'FM1* FM6* ',
            'SUS-ZM1_M2_COILOUTF_UL': 'FM1* FM6* ',
            'SUS-ZM1_M2_COILOUTF_UR': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_LL': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_LR': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_UL': 'FM1* FM6* ',
            'SUS-ZM3_M2_COILOUTF_UR': 'FM1* FM6* '}

 

This should have no effect on anything.

daniel.sigg@LIGO.ORG - 13:01, Tuesday 13 June 2023 (70417)

Remove the controls from these filter modules:

h1alsex {'LSC-X_ARM_DRIVE': 'FM3* FM5* '}
h1alsey {'LSC-Y_ARM_DRIVE': 'FM3* FM5* '}

Turned these filters off after a model restart.