Reports until 19:07, Sunday 11 June 2023
H1 General
austin.jennings@LIGO.ORG - posted 19:07, Sunday 11 June 2023 - last comment - 14:05, Tuesday 13 June 2023(70349)
Lockloss @ 2:02 UTC

LOCKLOSS @ 2:02, had a SRM saturation right before the lockloss. Seeing some movement in INP1 P and some LSC instability as well.

Images attached to this report
Comments related to this report
bricemichael.williams@LIGO.ORG - 14:05, Tuesday 13 June 2023 (70418)
Looking at the WIND and the SRM, the wind seems to gusting between 10-17mph at the same frequency as the oscillations in the SRM starting at 1m,21s before lockloss. 
Images attached to this comment