(Although the word "stabilized" is used loosely.)
Now that temperatures have seemed to settle in, attached* is a longer trend of the PR3, BS, and SR3 showing their new positions at these new LVEA temperatures. Note, this is a follow-on from Kissel's historical alog summary of the HVAC temperature changes 70284.
Recall that there were a few HVAC fan/setting changes with subsequent issues, failures, adjustments, and fire alarm testing interuptions which spanned the last month and rolled past just Tuesday maint periods. The PR3, BS, and SR3 seem to have slewed to new positions starting at the last LVEA temp change ~May 25th and have now settled in.
* V_INMON is the vertical channel for these suspensions - if the V channel shows change, it likely means that Pitch and or Yaw have also changed (PR3 pitch is shown to demonstarte). I confirmed this with previous plots, but chose to only plot this version since V tells the main story of epochs.
* Only 1 LVEA temp plot is shown for ease of viewing, but I checked via erlier versions of the plot that this zone temp change of over 1.5deg F is typical for what has been observed in the LVEA on other sensors.
We are still sorting new/old noise sources and any other such fallout from this new IFO alignment position.