kyle.ryan@LIGO.ORG - posted 16:14, Thursday 11 July 2013 (7051)
Pumped HAM 4 annulus
This volume had been vented until pumped this afternoon -> No change on the helium signal but there was a small response in total pressure indicating small inner O-ring leak on septum -> will leave under vacuum