Displaying report 1-1 of 1.
Reports until 16:54, Thursday 29 June 2023
H1 SQZ
naoki.aritomi@LIGO.ORG - posted 16:54, Thursday 29 June 2023 (70958)
NLG measurement

Naoki, Vicky

We measured NLG with seed. The measured NLG is 12.78, which corresponds to generated squeezing of 15.8 dB. The NLG calculator gives 12.64 so the calibration on February 3 is still OK.

Seed without pump: 0.00078
Seed with pump: 0.00997
NLG: 12.78

How to measure NLG

Displaying report 1-1 of 1.