TITLE: 07/02 Owl Shift: 07:00-15:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Corrective Maintanence
SHIFT SUMMARY:
- 10:22 - Went into Commissioning, CAMERA SERVO guardian went into WAIT FOR CAMERA due to the ASC CAM PIT/YAW 3 cameras freezing - sdfs attached
- Also noticed that the ETMY camera is now down, probably related since ETMY feeds in PIT/YAW 3 - ETMY camera does appear to be on and running, but showing up blue screnned on nuc26
- Tried letting the CAMERA SERVO guardian go through WAIT FOR CAMERA a few times before intervening
- Tried running through INIT to try and rerun through the code again - no dice - Tagging CDS
- IFO lost lock from EQ before I could find a solution
- 10:34 - inc 6.7 EQ from Tonga
- 10:39 EQ mode activtated
- Lockloss @ 10:40
- Holding in DOWN to ride out the ground motion
- INCREASE FLASHES failed 2x for ALS Y so I manually intervened during IA
- Same problem as Corey, I could not get PRC or SRC to align after multiple attempts trying to move BS and SRM
- Going to try skipping and going straight up ISC and see what happens
- Locking went surprisingly smooth - holding off at OMC WHITENING to damp violins
- Acquired NLN @ 14:15, but the ASC CAM PIT/YAW 3 on the CAMERA SERVO are still stuck due to ETMY cam failure and we cannot go into observing - will investigate
- On the camera process overview monit, I tried restarting the process, which failed, stating "Execution failed"
- Unfortunately not able to figure out how to restart the camera in time for the operator transition, though I believe it might be a quick fix by contacting Dave
LOG:
No log for this shift.