Summary of the report:
- The summary page is not updated after 10:00 UTC for 26th June 2023. The range plot is also not available after 06:00 UTC for 2nd July 2023.
- The average range throughout the week around 140 Mpc.
- Some of the locklossed during this week were due to ground motion, power glitches or maintenance. Some locklosses did not have any apparent reason.
- There is a constant high noise (~500 Hz) strain line present in the h(t) spectrogram on several days, except Wednesday and Thursday. No apparent cause found.
- The glitchgram showed an increased number of glitches on Saturday and Sunday. Mostly ground and wind motion were responsible for this.
- The noises in the BSCs got better during the end of the week.
- The EY accoustic showed low noises (~0-100 Hz) throughout the week, at different times of the day. Some of them can be attributed to ground motion and wind. Also, they might be the reason behind the glitches and the strain line noises.
- Hveto: H1:PEM-EX_VMON_ETMX_ESDPOWER48_DQ channel was the first round winner for everyday. The new channels are H1:LSC-REFL_A_LF_OUT_DQ, H1:SUS-ETMX_L3_OPLEV_PIT_OUT_DQ, H1:PEM-EX_VMON_ETMX_ESDPOWER18_DQ. The third channel had appeared in the previous week.
- Fscan: All the plots have appeared. Nothing significantly new.
Link to the full report: Report