david.barker@LIGO.ORG - posted 08:22, Sunday 09 July 2023 - last comment - 09:39, Sunday 09 July 2023(71174)
Restarted h1isiitmx, part of lock-loss investigation

Jim, Ryan, Austin, Dave:

We restarted the h1isiitmx model as part of the lock-loss investigation.

Sun09Jul2023
LOC TIME HOSTNAME     MODEL/REBOOT
08:13:50 h1seib3      h1isiitmx   
 

Comments related to this report
david.barker@LIGO.ORG - 08:49, Sunday 09 July 2023 (71175)

model restart did not fix anything, we are now power cycling the IO Chassis

david.barker@LIGO.ORG - 09:13, Sunday 09 July 2023 (71176)

IO Chassis power cycle did not fix anything. Austin reports all the B3 chassis externally look good (green leds).

david.barker@LIGO.ORG - 09:21, Sunday 09 July 2023 (71178)

Sun09Jul2023
LOC TIME HOSTNAME     MODEL/REBOOT
08:13:50 h1seib3      h1isiitmx   
08:56:51 h1seib3      ***REBOOT***
08:58:24 h1seib3      h1iopseib3  
08:58:37 h1seib3      h1isiitmx   
08:58:50 h1seib3      h1hpiitmx   
 

david.barker@LIGO.ORG - 09:39, Sunday 09 July 2023 (71179)

SEI is heading to the site to investigate