I updated the swept-sine frequency vectors used in the calibration swept-sine DTT templates. They now match what is used in LHO's simulines configuration.
Frequency points changed (Hz):
10.01 -> 10.11
59.78 -> 64.78
180.5 -> 175.5
DTT templates are located at /ligo/groups/cal/H1/ifo/templates. pyDARM cmd tools will automatically use these templates on future runs.
Reasons behind changes were to avoid the QUAD bounce mode; and steer clear of the 60 Hz line and its harmonics. These impacted the transfer function and coherence of the measurements.