Reports until 11:58, Thursday 13 July 2023
H1 CAL
louis.dartez@LIGO.ORG - posted 11:58, Thursday 13 July 2023 - last comment - 17:28, Thursday 13 July 2023(71298)
updated calibration DTT templates
I updated the swept-sine frequency vectors used in the calibration swept-sine DTT templates. They now match what is used in LHO's simulines configuration.

Frequency points changed (Hz): 

10.01 -> 10.11
59.78 -> 64.78
180.5 -> 175.5

DTT templates are located at /ligo/groups/cal/H1/ifo/templates. pyDARM cmd tools will automatically use these templates on future runs.
Comments related to this report
vladimir.bossilkov@LIGO.ORG - 15:52, Thursday 13 July 2023 (71303)

Reasons behind changes were to avoid the QUAD bounce mode; and steer clear of the 60 Hz line and its harmonics. These impacted the transfer function and coherence of the measurements.

louis.dartez@LIGO.ORG - 17:28, Thursday 13 July 2023 (71310)
I'll add that we're building the confidence to move completely over to simulines for future calibration injection suites. However, it's good to have the corresponding DTT templates kept up to date for the scenario in which they're needed.