Comparing the new ramped fill, where the LLCV is opened linearly from its operational value to 100% over 5 minutes, with the unramped opening (takes about 5 seconds) the main difference is the removal of the sudden jump in discharge line pressure and temperatures at the start of the fill.
Log and linear comparison plots are attached showing today's and Sunday's fills.